Sample XY Stage

SampleXY.jpg
Vertical stage is xafs_liny (alias for XF:06BMA-BI{XAFS-Ax:LinY}Mtr). Limits are [-1:485].

Horizontal stage is xafs_linx (alias for XF:06BMA-BI{XAFS-Ax:LinX}Mtr).  Limits are [140:-150]

Ion Chambers and electrometer

Voltage on ion chambers is 800 V. The 500 to 600 range seems to be in the tail end of the ohmic response.  700 V to 900 V seems pretty constant with constant flux.

Electrometer from Pete Siddons' group, used to read ion chambers.
read signal from channel 1: caget XF:06BM-BI{EM:1}EM180:Current1:MeanValue_RBV

stop: caput XF:06BM-BI{EM:1}EM180:Acquire 0

start: caput XF:06BM-BI{EM:1}EM180:Acquire 1

Set a tenth of second: caput XF:06BM-BI{EM:1}EM180:AveragingTime 0.1

With the sampling time set at 4 miliseconds, a 0.1 second averaging time accumulates and averages 25 measurements.  Similarly, a 0.5 second averaging time accumulates 125 measurements.

Reference foils

Linxs.jpg

Reference foils are mounted on stage xafs_linxs (alias for XF:06BMA-BI{XAFS-Ax:LinXS}Mtr) , thus can be programmed into a measurement macro.  The top-most foil is at position xafs_linxs = -90, the bottom-most foil is at +90.

Motor parameters


VMAX
VELO
MRES
DIR
OFF
HLM
LLM
aka
linx
4
4
0.0003175
Neg
-60



liny
4
4
0.0003175
Pos
-25



lins
4
4
0.0003175
Pos
0



linxs
4
4
0.000125
Neg
18



pitch
0.25
0.25
0.000078125
Pos
0.03
10
-10

roll
0.25
0.25
0.000078125
Pos
-0.14
10
-10

roth1
1
0.5
3.90625e-06
Pos
-4.6
65.4
-64.6

rotb2
5
5
0.000390625
Neg
46.532
7200
-7200
wheel
ref
15
15
0.0003901
Neg 296.4993435
7200
-7200

rots3
5
4
0.000351748
Neg
15.9325
3600
-3600

tbl_yu
2
2
1.921875E-5
Pos
-64



tbl_ydi
2
2
1.921875E-5
Pos
6



tbl_ydo
2
2
1.921875E-5
Pos
-17



tbl_xu
1
1
5E-5
Neg
-21.2



tbl_xd
1
1
5E-5
Neg
23.4